Image Enhancement Technology
Example - DF femoral artery
ROI centred on femoral artery.
 |
Original DF image + Region Of Interest (ROI) |
 |
7x ROI before enhancement |
 |
7x ROI before enhancement with edges marked |
 |
7x ROI after enhancement with edge-dots removed |
 |
7x ROI after enhancement with edges marked and measurements indicated
(for measurements, see below) |
The absolute measurements via the calibration by a standard wire are:
- Diameter of .035J wire = 2.29 ± .02 pixels (=> 0.89mm)
- Normal patient vessel lumen diameter = 13.29 ± .02 pixels => 5.17 ± .01mm
- Patient vessel within stenosis cleared of blockage = 5.29 ± .02 pixels => 2.06 ± .01mm
- Stenosed vessel lumen diameter = (13.29 - 5.29) ± .04pixels => 3.11 ± .02mm
- Percentage diameter stenosis of superficial femoral artery = 3.11/5.17 => 60.15 ± 0.84%
Before and after enhancement, side-by-side:
'before' |
'after' |
 |
 |